Colonoscopy can be associated with many problems, such as mechanical trauma due to the distal tip contacting the colon wall or health issues due to the extended use of anesthesia. In order to eliminate these complications, an automatic adjustable colonoscope was designed. This device uses sensors, actuators, and a control system to automatically position the distal tip in the center of the colon lumen. The sensors were tested to determine their ability to accurately sense the distance from the tip to the surface. The actuators were tested to determine the correlation between motor rotation and displacement of the distal tip. The control system was tested to assess the ability of the device to position the tip in the center of the test tube and the ability to navigate through a flat test course. It was determined that the sensors could accurately determine distances from 0 to 15 mm from the test surface in all test conditions. The motors for up-down movement and left-right movement of the colonoscope had response times of 0.57 s and 0.69 s, respectively, when the motors were rotated from 0 deg to 90 deg. The control system was able to safely move the colonoscope tip away from all walls of the test apparatus. It was also able to navigate through the flat test course without coming in contact with the walls. The automatic adjustable colonoscope has demonstrated that it can safely and effectively position the distal tip to avoid contact with the walls of the test surface.
Skip Nav Destination
e-mail: choij1@ohio.edu
Article navigation
December 2011
Technical Briefs
Development of an Automatic Adjustable Colonoscope
Jonathan D. Litten,
Jonathan D. Litten
Department of Mechanical Engineering,
Ohio University
, 259 Stocker Center, Athens, OH 45701
Search for other works by this author on:
JungHun Choi,
JungHun Choi
Department of Mechanical Engineering and Biomedical Engineering Program,
e-mail: choij1@ohio.edu
Ohio University
, 254 Stocker Center, Athens, OH 45701
Search for other works by this author on:
David Drozek
David Drozek
College of Osteopathic Medicine, Department of Specialty Medicine,
Ohio University
, 106 Parks Hall, Athens, OH 45701
Search for other works by this author on:
Jonathan D. Litten
Department of Mechanical Engineering,
Ohio University
, 259 Stocker Center, Athens, OH 45701
JungHun Choi
Department of Mechanical Engineering and Biomedical Engineering Program,
Ohio University
, 254 Stocker Center, Athens, OH 45701e-mail: choij1@ohio.edu
David Drozek
College of Osteopathic Medicine, Department of Specialty Medicine,
Ohio University
, 106 Parks Hall, Athens, OH 45701J. Med. Devices. Dec 2011, 5(4): 044502 (6 pages)
Published Online: November 14, 2011
Article history
Received:
August 3, 2011
Revised:
October 20, 2011
Online:
November 14, 2011
Published:
November 14, 2011
Citation
Litten, J. D., Choi, J., and Drozek, D. (November 14, 2011). "Development of an Automatic Adjustable Colonoscope." ASME. J. Med. Devices. December 2011; 5(4): 044502. https://doi.org/10.1115/1.4005379
Download citation file:
Get Email Alerts
Cited By
Related Articles
Development and Performance of a Controllable Autoloading Needle-Free Jet Injector
J. Med. Devices (March,2011)
Design of a Bending Section Auto-Driven Mechanism for Colonoscope
J. Med. Devices (June,2009)
Experiments on Active Vibration Control of a Flexible Four-Bar Linkage Mechanism
J. Vib. Acoust (January,2000)
Parametric Design of
Parallel Force/Velocity Actuators: Force Distribution
Analysis
J. Mechanisms Robotics (February,2010)
Related Proceedings Papers
Related Chapters
QP Based Encoder Feedback Control
Robot Manipulator Redundancy Resolution
Managing Energy Resources from within the Corporate Information Technology System
Industrial Energy Systems
Accuracy of an Axis
Mechanics of Accuracy in Engineering Design of Machines and Robots Volume I: Nominal Functioning and Geometric Accuracy