The Thermal Pulse Decay (TPD) method for the determination of local tissue thermal conductivity and blood perfusion rate is based on a comparison of measured with theoretically calculated temperatures. A sensitivity analysis of the theoretical model is performed. This analysis supports the establishment of an experimental protocol which reduces the measurement errors: An “optimal” measurement time inverval for typical perfusion rates (up to 6 mL/mL/min) was found to be between 3 and 11 s after the heat pulse is turned off. Within this interval, the maximum error in determination of tissue conductivity and blood perfusion caused by experimental measurement errors is expected not to exceed 5 percent. The presently chosen pulse duration of 3 s is in agreement with the analysis as a good compromise between accuracy and excessive tissue heating.
Skip Nav Destination
Article navigation
February 1986
Research Papers
A Sensitivity Analysis of the Thermal Pulse Decay Method for Measurement of Local Tissue Conductivity and Blood Perfusion
H. Arkin,
H. Arkin
Department of Mechanical and Industrial Engineering, and Department of Veterinary Biosciences, University of Illinois at Urbana-Champaign, Urbana, Ill. 61801
Search for other works by this author on:
K. R. Holmes,
K. R. Holmes
Department of Mechanical and Industrial Engineering, and Department of Veterinary Biosciences, University of Illinois at Urbana-Champaign, Urbana, Ill. 61801
Search for other works by this author on:
M. M. Chen
M. M. Chen
Department of Mechanical and Industrial Engineering, and Department of Veterinary Biosciences, University of Illinois at Urbana-Champaign, Urbana, Ill. 61801
Search for other works by this author on:
H. Arkin
Department of Mechanical and Industrial Engineering, and Department of Veterinary Biosciences, University of Illinois at Urbana-Champaign, Urbana, Ill. 61801
K. R. Holmes
Department of Mechanical and Industrial Engineering, and Department of Veterinary Biosciences, University of Illinois at Urbana-Champaign, Urbana, Ill. 61801
M. M. Chen
Department of Mechanical and Industrial Engineering, and Department of Veterinary Biosciences, University of Illinois at Urbana-Champaign, Urbana, Ill. 61801
J Biomech Eng. Feb 1986, 108(1): 54-58 (5 pages)
Published Online: February 1, 1986
Article history
Received:
October 6, 1984
Revised:
August 26, 1985
Online:
June 12, 2009
Citation
Arkin, H., Holmes, K. R., and Chen, M. M. (February 1, 1986). "A Sensitivity Analysis of the Thermal Pulse Decay Method for Measurement of Local Tissue Conductivity and Blood Perfusion." ASME. J Biomech Eng. February 1986; 108(1): 54–58. https://doi.org/10.1115/1.3138580
Download citation file:
Get Email Alerts
Related Articles
A Spherical Source Model for the Thermal Pulse Decay Method of Measuring Blood Perfusion: A Sensitivity Analysis
J Biomech Eng (February,1989)
Thermal Pulse Decay Method for Simultaneous Measurement of Local Thermal Conductivity and Blood Perfusion: A Theoretical Analysis
J Biomech Eng (August,1986)
A Step-Wise Method for Measuring Thermophysical Parameters of an Isotropic Material, Considering the Effect of Thermal Contact Resistance and Lateral Convection Losses
J. Heat Mass Transfer (January,2025)
Analysis
of Bioheat Transport Through a Dual Layer Biological
Media
J. Heat Transfer (March,2010)
Related Proceedings Papers
Related Chapters
Experimental Investigation of an Improved Thermal Response Test Equipment for Ground Source Heat Pump Systems
Inaugural US-EU-China Thermophysics Conference-Renewable Energy 2009 (UECTC 2009 Proceedings)
Tissue and blood-material interactions
Biocompatible Nanomaterials for Targeted and Controlled Delivery of Biomacromolecules
Scope of Section I, Organization, and Service Limits
Power Boilers: A Guide to the Section I of the ASME Boiler and Pressure Vessel Code, Second Edition