Recent applications of 20 MHz pulsed ultrasound Doppler velocimetry (PUDVM) in microsurgical research have necessarily employed piezoelectric crystals whose diameter is not negligible compared to the lumen size (1–2 mm) of many vessels of interest. A three-dimensional numerical model was developed to explore relationships between actual and detected flow field parameters, for (steady) Poiseuille flow, when appreciable velocity gradients exist within the PUDVM sample volume. Validation studies showed that highly accurate velocity profiles could be obtained in the limiting case of a very small sample volume (0.1 mm radius), but that for currently employed crystals ( ≈ 0.5 mm radius) there was appreciable underestimation of the centersteam velocity, and appreciable overestimation of the flow stream diameter. Errors in perceived velocity and flow rate were found to be relatively insensitive to perturbations in the sample volume thickness, in the size of the sampling range increment, or in the angle of insonation beam divergence. By contrast, these apparent flow parameters were found to be very sensitive to perturbations of sample volume diameter or of the Dopper angle. Small variations in the degree of partial sample volume overlap of the flowstream periphery were shown to be capable of causing large fluctuations in apparent flow stream diameter.
Skip Nav Destination
Article navigation
August 1985
Research Papers
A Three-Dimensional Computational Simulation of Some Sources of Measurement Artifact in Microvascular Pulsed Ultrasound Doppler Velocimetry
T. D. Brown,
T. D. Brown
Departments of Orthopaedic Surgery and Biomedical Engineering, The University of Iowa, Iowa City, Iowa 52242
Search for other works by this author on:
L. D. Bell,
L. D. Bell
Departments of Orthopaedic Surgery and Biomedical Engineering, The University of Iowa, Iowa City, Iowa 52242
Search for other works by this author on:
D. R. Pedersen,
D. R. Pedersen
Departments of Orthopaedic Surgery and Biomedical Engineering, The University of Iowa, Iowa City, Iowa 52242
Search for other works by this author on:
W. F. Blair
W. F. Blair
Departments of Orthopaedic Surgery and Biomedical Engineering, The University of Iowa, Iowa City, Iowa 52242
Search for other works by this author on:
T. D. Brown
Departments of Orthopaedic Surgery and Biomedical Engineering, The University of Iowa, Iowa City, Iowa 52242
L. D. Bell
Departments of Orthopaedic Surgery and Biomedical Engineering, The University of Iowa, Iowa City, Iowa 52242
D. R. Pedersen
Departments of Orthopaedic Surgery and Biomedical Engineering, The University of Iowa, Iowa City, Iowa 52242
W. F. Blair
Departments of Orthopaedic Surgery and Biomedical Engineering, The University of Iowa, Iowa City, Iowa 52242
J Biomech Eng. Aug 1985, 107(3): 274-280 (7 pages)
Published Online: August 1, 1985
Article history
Received:
September 22, 1984
Revised:
May 16, 1985
Online:
June 15, 2009
Article
Article discussed|
View article
Connected Content
Citation
Brown, T. D., Bell, L. D., Pedersen, D. R., and Blair, W. F. (August 1, 1985). "A Three-Dimensional Computational Simulation of Some Sources of Measurement Artifact in Microvascular Pulsed Ultrasound Doppler Velocimetry." ASME. J Biomech Eng. August 1985; 107(3): 274–280. https://doi.org/10.1115/1.3138553
Download citation file:
Get Email Alerts
Related Articles
A First-Order Mechanical Device to Model Traumatized Craniovascular Biodynamics
J. Med. Devices (March,2007)
A Longitudinal Study of Remodeling in a Revised Peripheral Artery Bypass Graft Using 3D Ultrasound Imaging and Computational Hemodynamics
J Biomech Eng (April,2011)
The Effect of the Accommodation Coefficient on Slider Air Bearing Simulation
J. Tribol (April,2000)
The Simulation of Discrete Vessel Effects in Experimental Hyperthermia
J Biomech Eng (August,1994)
Related Proceedings Papers
Related Chapters
A Nonlinear Power Control Strategy for Ultra-High Voltage Tie Line
International Conference on Electronics, Information and Communication Engineering (EICE 2012)
Numerical Simulation of High Intensity Focused Ultrasound (HIFU) Using a Fully Compressible Multiscale Model
Proceedings of the 10th International Symposium on Cavitation (CAV2018)
Numerical Simulation of the Bubble Cloud Dynamics in an Ultrasound Field
Proceedings of the 10th International Symposium on Cavitation (CAV2018)